4-78
Maintenance Procedures
Replacing the IonSABRE II probe sample capillary
Replace the stainless steel sample capillary in the IonSABRE II probe if it
becomes blocked and you cannot clear it, or if it becomes contaminated or
damaged.
Removing the existing capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm wrench
•
Combined 2.5-mm Allen wrench and cone extraction tool
To remove the existing capillary:
1. Remove the probe from the source (see
).
2. Retrieve the combined 2.5-mm Allen wrench and cone extraction tool
from its storage location on the source adaptor housing.
Warning:
The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The probe and source can be hot. To avoid burn injuries, take
great care while performing this procedure.
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...