4. Load and run the method-screening sample list.
Tip:
During the scouting run, after you identify the optimum conditions for
the run, to quit the remainder of the sample list, in the Main window, click
"Skip to the next sample list"
. In the event that the queue is empty, the
system immediately switches to system-shutdown conditions.
5. Open and view the data-log file generated from the scouting run.
6. Analyze the data from the scouting run, and use it to confirm the optimum
conditions for separating your fractions.
Analyze the data from a scouting run:
For more information, see these topics in the ChromScope online Help:
l
Create sample lists for acquiring data and collecting fractions
l
Create sample lists using the Sample List wizard
l
Work with integration
22
August 8, 2014, 715003940 Rev. B
Prep 150 LC System