![WaterLink SpintouchDW 3585 Manual Download Page 10](http://html1.mh-extra.com/html/waterlink/spintouchdw-3585/spintouchdw-3585_manual_3621849010.webp)
10
Certifications
EZ-BLE
™
PRoC
™
Module,
CYBLE-022001-00 RF
Radio:
FCC (USA):
Industry
Canada (IC)
Certification:
FCC ID: WAP2001
License IC: 7922A-2001
CE (Europe):
Complies with Directive 1999/5/EC
MIC (Japan):
005-101007
KC (Korea):
MSIP-CRM-Cyp-2001
EMC:
EU: ETSI EN 301489-1; US: FCC PART 15 B; CAN ICES-3
(B)/NMB-3(B); AS/NZS: CSPR 22
Safety:
EU: EN61010-1:2010; AS/NZS: national differences
Dimensions
21.6 X 12.4 X 10.4 cm (L X W X H) | 8.5 X 4.9 X 4.2 in
Weight
0.79 Kg, 1.74 lb
TROUBLESHOOTING
TROUBLESHOOTING GUIDE
Problem
Reason
Solution
Meter Check Disk
Meter Check Disk in chamber
instead of reagent disk
Select “Continue” to go to Test Results
screen. Select “Abort” to go to Testing
screen and run test with reagent disk.
“Sanitizer” appears
on test screen
instead of “Source”
Wrong Market Type was
chosen
Go to Settings. Change Market Type to
Drinking Water
!
On Test Results
and Test History
screens
Problem with default blank
due to under filled disk or
air bubble. Test results are
questionable.
Fill disk correctly (see FILLING). On Test
Results screen tap ! for details.
Range Error
Raw data out of range
Contact Support
Output Error
Decreased light intensity.
Possibly dirty lens
Clean lens (see CLEANING). Follow
Range Check Procedure. If error
message persists, contact Support.
Consistently
unexpected high
results for metals
Metals may actually be
present
Repeat test with distilled water. If
the results still show that metals are
present, contact Support.
Alkalinity result of 0
ppm
Usually due to an underfilled
disk.
Review the disk filling procedures and
test again. If problem persists, contact
Support.
Unexpected results
Dirty disk cover
Gently clean disk cover apertures with
pipe cleaner or lint free cloth.
Disk type is not an
option in Disk Series
Software or meter firmware is
out of date.
Update WaterLink Connect 2 at
softwarecenter.lamotte.com
High pH results
Water sample temperature
above 100 °F (38 °C)
interferes with pH reagent
For water samples over 100 °F (38 °C)
subtract 0.3 from pH result or, for the
most accurate result, wait until water
sample is below 90 °F (32 °C) to test.
Disk not spinning
Lid open, meter not powered
on, low battery, disk or disk
cover pressed down too tightly
on hub
Close lid, power on meter, charge the
battery or plug meter into a stable
power source, remove the disk/disk
cover and place back in the meter
more gently.
Fast electrical transients may
disrupt operation of the Spin
Touch meter
Restart the test to resume normal
operation.
Test result value on
the display is red
Results are out of range of
reagent system
Dilute sample. Retest for out of range
test factor. (All factors except pH.)