8
¬
Gently wipe smudges from the screen with the Cloth Wipe (3580-WIPE).
¬
Do not touch the screen with a sharp object.
¬
Do not place objects on the screen that will scratch or damage it.
¬
Avoid touching the screen with wet fingers.
TESTING
12:14 AM
16-Apr-20
Sani
ti
zer
Disk Series
Salt
203
Tap to
Start Test
1. Press
and hold until the meter turns on.
2. Tap . Select a sanitizer type. Tap to confirm.
3. Tap . Select a disk series (found on disk packaging). Tap
to confirm.
NOTE: Disk Series are limited by Sanitizer selection.
4. Remove a disk from the packaging.
5. Use the syringe (1189) to fill the disk with the pool or spa water sample.
6. Insert the disk. Cover the disk with the Universal Disk Cover (1719). Close the lid.
7. Tap
to start test. Tap
to cancel the test. If the test is cancelled discard the
disk.
8. The results will be displayed.
9. Choose an option.
¬
Tap
to add a tag.
¬
Tap the highlighted
to save the test results to the test log if Auto Save is
not enabled.
¬
Tap the hightlighted
to send the results to the enabled Mobile Bluetooth
Printer.
¬
Tap the highlighted
to send the results to a Bluetooth enabled device.
¬
Tap
to return to the Test Screen.
10. Press and hold
for 2 seconds to turn the meter off.
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the most accurate
result, wait until water sample is below 90 °F (32 °C) to test.
Remove salt residue daily. Salt will damage the meter and cause inaccurate results. See cleaning page
20.
GENERAL OPERA
TING PROCEDURES