Warning

• 

Never operate the monitor when the cover is removed.

• 

Remove the monitor cover and battery only in area known as 

non-hazardous.

• 

Use only WatchGas lithium battery part number M500-0001-

000 [1.17.02.0002] (3.6 V, 2700 mAh, AA size).

• 

This instrument has not been tested in an explosive gas/air 

atmosphere having oxygen concentration greater than 21%.

• 

Substitution of components will impair suitability for intrinsic 

safety and void warranty.

• 

It is recommended to bump test with a known concentration 

gas to confirm the instrument is functioning properly before 
use.

• 

Before use, ensure that the colorless ESD layer on the display 

is not damaged or peeling. (The blue protective film may be 

removed.)

Avertissement

• 

N’utilisez jamais le moniteur lorsque le couvercle est enlevé.

• 

Retirer le couvercle du moniteur et la batterie uniquement 

dans une zone connue comme non dangereuse.

• 

Utilisez uniquement le numéro de pièce de la batterie au 

lithium mPower M500-0001-000 [1.17.02.0002].

• 

Cet instrument n’a pas été testé dans une atmosphère explosive 

gaz / air ayant une concentration en oxygène supérieure à 21%.

• 

La substitution de composants compromettra l’aptitude à la 

sécurité intrinsèque et annulera la garantie.

• 

Il est recommandé de tester avec un gaz de concentration 

connu pour confirmer que l’instrument fonctionne 

correctement avant de l’utiliser.

• 

Avant l’utilisation, assurez-vous que la couche ESD incolore de 

l’écran n’est pas endommagée ou épluchée. (Le film protecteur 

bleu peut être enlevé.)

Read Before Operating

The User’s Guide must be carefully read by all individuals who have 

or will have the responsibility of using, maintaining, or servicing 

this product. The product will perform as designed only if it is used, 

maintained, and serviced in accordance with the manufacturer’s 

instructions.

Summary of Contents for UNI Sustainable

Page 1: ...Quick Start Guide D e t e c t i o n UNI Sustainable Single Gas Detector...

Page 2: ...le du moniteur et la batterie uniquement dans une zone connue comme non dangereuse Utilisez uniquement le num ro de pi ce de la batterie au lithium mPower M500 0001 000 1 17 02 0002 Cet instrument n a...

Page 3: ...alarm buzzer two keys Left Key Confirm Number increasing and Right Key Power Cursor moving an alligator clip and a chemical sensor ALLIGATOR CLIP BATTERY STATUS SENSOR WITH ID ON LABEL POWER CURSOR M...

Page 4: ...n mark perform operation 3 OK status indicator or accept number entry 4 Concentration unit e g ppm mg m3 etc 5 Battery charge status 6 Alarm type indicator 7 Zero in progress indicator 8 Calibration i...

Page 5: ...sec count down until the unit powers off Normal mode In normal mode the gas concentration displayed on the LCD continuously and the unit alarms if a pre set limit is exceeded The user can check vario...

Page 6: ...ON ZERO SET Power on zero enable disable FAST SET Fast power on enable disable RST CF9 Reset configuration to default values EXIT Exit Config Mode In Config Mode in general use the Left Key to incre a...

Page 7: ...s the base line for the sensor and is done in fresh air or other clean air source Enter Config Mode and AiR displays as the first menu item Press the Left key to start the15 second zero calibration co...

Page 8: ...e g hydrogen chloride hydrogen fluoride ozone chlorine dioxide phosphine ammonia and ethylene oxide be sure to use Teflon tubing connections 2 Enter Config Mode and scroll to SPAN 3 Start the gas flo...

Page 9: ...t filter regularly Battery and sensor can be replaced as needed For other maintenance contact mPower s service department or refer to the full User Manual REMOVE 4 SCREWS REPLACE SENSOR REPLACE BATTER...

Page 10: ...This product may contain one or more nickel metal hydride NiMH lithium ion or alkaline batteries Specific battery information is given in this user guide Batteries must be recycled or disposed of pro...