background image

 

 
 

 
3.2 Decimal point setting 
  Prior to factory shipment, the decimal point of this meter is so 

set that its position corresponding to the 10

1

 digit is lit, change 

the position of the decimal point by referring to the following 
figure. 

 
 
 
 
 
 
 
 
 
 
 

4. Calibration 

 In order to calibrate the meter, a current generator with an accuracy 

of ±0.01% or more is required. This calibration is conducted in 
the same manner as 3.1 Scaling adjustment. 

 

5. Specifications 

Model 

Measuring 

range 

Display 

Max. allowable 

input current 

AL-512 

4 to 20mA DC 

Offset;±200 

Fullscale; 

100 to 1999 

±50mA 

Accuracy: ±(0.1% of FS +1 digit)(23℃±5℃ and 35 to 85% RH) 

   

Measurement function 

: Measurement  of  instrumentation 

signal(4 to 20mA DC) 

Operation  

:Double integral type 

Maximum display  

:1999 

Sampling speed 

:2.5 times/sec.(TYP) 

Noise rejection ratio  :NMR 40dB (TYP) 
Temperature 
characteristics 

:Offset displayed-value: 
  ±0.3 digit/℃ 

Fullscale displayed-value: 
±0.3 digit/℃ 

Overrange warning 

:For an input of more than 1999, “1” 

is displayed in the most 
significant digit and the lower 3 
digits go out. 

Variable span of 
fullscale  

:100 to 1999 

Variable span of 
offset 

:±200 

Voltage drop 

:5V at an input of 4mA(TYP) 
  5.3V at an input of 20mA(TYP) 

Display 

:7-segment LEDs (Red light emitting 

diode numeric elements) 

  Character height, 14.2mm 

Polarity display 

:”-“ is automatically displayed 

when the computation result is 
minus. 

External control 

:None 

Operating temperature 
/humidity range 

:0 to 50℃/ 35 to 85%RH  

Power supply 

:None (Driven by input signal) 

Dielectric  strength 

:Between input terminal (LO) and E 

terminal; For 1 min. at 500V DC 

Dimensions 

:96(W)×48(H)×22(D)mm 

Weight 

:Approx 55g 

Accessory 

:Screw terminal cover 1 pc. 

Instruction manual 1 copy 

Conformity standard 

:EN61326-1: 2006 

 
 
 
 
 
 

 

6. Dimensions 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

7. Warranty and After Sales Service 

7.1 Warranty 

The warranty lasts for one year from the date of delivery. If this 
product fails during this period and the reason is considered to 
be clearly.        
The  manufacturer  warrants  to  the  original  retail  customer  its 
indicator to be free of defects in material and workmanship for 
use under normal care and will repair or replace any 

7.2 After Sales Service 

Under  strict  quality  control  measures,  this  product  was 
manufactured, tested, inspected and shipped. Should a defect in 
manufacture  or  Workmanship  be  identified,  please  return  the 
product to our distributor or directly to us. It would be highly 
appreciated if you could give a detailed account of the fault and 
enclose it with the product. 

 
 
 

Model AL-512 Series                                                                          ED-49982c 

WATANABE ELECTRIC INDUSTRY CO., LTD.

6-16-19, Jingumae, Shibuya-ku, Tokyo 150-0001, Japan 
Phone: (81)3-3400-6140
Homepage http://en.watanabe-electric.co.jp/

ASAHI KEIKI DIVISION

WATANABE ELECTRIC INDUSTRY CO., LTD.

6-16-19, Jingumae, Shibuya-ku, Tokyo 150-0001, Japan 
Phone: (81)3-3400-6141

Homepage http://www.watanabe-electric.co.jp/en/