9 Troubleshooting
Fault
Cause
Solution
No display reading
and/or only ON/OFF
switch illuminated
Blown fuse
Replace fuse and trace cause or contact
Service Department
Temperature/pres
sure too high
Check pressure gauge – if pressure in red
range (above 6.3 bar), check for fault and
repair if necessary (Service Workshop)
Hot water rather than
steam comes from
nozzle
Water container
too full
Set unit to dry steam and spray until only
steam comes out
Pressure gauge below 0
bar
Normal function
None
Collector spills over
Full
Empty collector
Wet steam setting
produces only dry steam
Control turned to
“0“
Select wet steam (settings 1-5)
Low-water LED on
continuously
No water flow
Check tap and connections
Low-water LED flashes
in conjunction with wet-
steam and dry-steam
LEDs
Indicates need to
flush container
Flush container as instructed
Water-supply LED on
continuously
No water supply
Check tap and connections
Heating LED on
continuously with no
increase in pressure
Faulty heater
Contact Service Department
Water-leak LED comes
on
Unit leaking
Contact Service Department
If the above recommendations do not solve the problem, contact your dental depot
or our service department.