9 Troubleshooting
Fault
Cause
Solution
Polyquick 260
does not heat
Lid is open (Fig. 1)
Close lid against safety contact.
Temperatur-programme-
parameter is to low
Adapt the parameters.
Preheating temperature
is to low
Adapt the parameters.
Heater faulty
Please contact our service department.
Pressure does
not increase in
chamber
Lid seal O-ring faulty
Check O-ring and replace if necessary.
Pressure regulator set to 0
(pressure gauge).
Check pressure and adjust if necessary.
Lid not fully closed
Close lid against safety contact or
contact faulty.
Temperature monitor off or faulty Press in safety knob (left side of unit)
and/or replace knob. Please contact our
service department.
Display blank
Temperature monitor off or faulty Press in safety knob (left side of unit)
and/or replace knob. Please contact our
service department.
Touchscreen
malfunction
Calibrate the touch display
Switch on the unit. While the logo of
Wassermann appears please contact
the touch display two times.
Then you will reach the self-explanatory
touch panel-calibration (Fig. 2-3).
Touchscreen under pressure
Please contact our service department.
Screen protector damaged?
Please contact our service department.
Unit does
nothing
Blown fuse
Replace fuse at rear of unit.
If the above recommendations do not solve the problem, contact your dental
depot or our service department.