4
Specification
The MR200 is MR Conditional to 7.0T for the parts installed inside the MR scan room:
Audio Control Unit -
MR Unsafe
Microphone -
MR Unsafe
DVD/CD Player -
MR Unsafe
Noise Reducing Headphones -
MR Conditional to 7T
Earphones -
MR Conditional to 7T
Stereo Acoustic Converter -
MR Conditional to 7T
Important note: Equipment should be set up in accordance with the layout diagram
(page 15).
MR Unsafe items must not be taken into the MR scan room.
Width
mm
250
Depth
mm
300
Height
mm
100
Weight
Kg
2.5
Width
mm
190
Depth
mm
190
Height (max)
mm
500
Weight
Kg
1
Width
mm
120
Depth
mm
170
Height
mm
500
Weight
Kg
0.5
Audio Control Unit
Microphone
Stereo Acoustic Convertor