> Clean the handpiece with cable
immediately after every treatment, to flush out liquid (e.g., blood, saliva etc.) and to
prevent settling on the internal parts.
> Operate the coolant fill function for at least 10 seconds.
> Ensure that all coolant outlets are rinsed out.
> Remove the instrument.
> Remove the handpiece with cable.
> Wipe the entire surface of the handpiece with cable, the universal support and the stand with disinfectant.
Note that the disinfectant used during pre-treatment is only for personal protection and cannot replace the disinfecting
step after cleaning.
Hygiene and maintenance
Initial treatment at the point of use