16
Start-up
General
The built-in air filter in the connecting hose is only a fine filter.
It cannot clean unfiltered and possibly contaminated compressor air.
Dirty fine filters must be replaced.
> Use only non-contaminated,oil-free,dry and filtered compressed air according to ISO 7494-2.
Fit the suction filter so that it
is leak-tight.
Push the quick release coupling of
the connecting hose into the
coupling socket on the Assistina.
Connect the compressed air line.
02697000