9072 Hydrazine Analyzer
Method 2:
Prepare a solution with a defined hydrazine concentration in 5 liter canisters.
Example: Approach of the dilution series
a) Starting solution 24% hydrazine hydroxide (H
2
H
5
OH) = 15% N
2
H
4
= 150 g/ l N
2
H
4
b)
Solution A:
Add10 ml starting to 1000 ml deionized water = 15 g/ l N
2
H
4
b)
Solution B:
Add 10 ml
solution A
to 1000 ml deionized water = 150 g/ l N
2
H
4
Calibration solution:
Add 16.7 ml of
solution B
to 5 l hydrazine-free water = 500μg/ l N
2
H
4
You start the calibration procedure by pressing key
and insert your PIN number.
Select SubCal:
The setting of the present hydrazine concentration is performed digit for digit. To reach the
measuring mode, press the key ESC twice.
If the value is beneath 10 % of the value determined through the calibration, which followed
after a cleaning procedure, a new cleaning of the electrode is necessary.
SubCal
000.0 μg/l