START-UP
SubCal >
you are setting the present hydrazine concentration digit for digit, finally push
key SET.
3
rd
Menu
level
Range Modus >
gives you the choice between
ManRange
(range which is
setting in 1
st
level only) and
AutoRange
(automatic changing range which has double
value of
ManRange
>)(Option only).
4
th
Menu
level
I-Output
> sets the analog output current 0(4)...20 mA.
5
th
Menu
level
Limit
> allows limit settings. There exists a switch hysteresis of 10% of
the chosen limit.
6
th
Menu
level
FlowCal
> enables flow meter calibration (see
Section 7
).
7
th
Menu
level
Parameter
> shows operation parameters Kd (flow) and k (calibration).
Attention
:
Only reading parameter.
8
th
Menu
level
Logging
> is the data logging function (Option only).
9
th
Menu
level
Date/Time
> allows date settings.
10
th
Menu
level
Service >
is for service only.
11
th
Menu
level
ChangePIN
> gives you the possibility to change your PIN number.
Attention: lost PIN numbers can only activated by the manufacturer.