18
Calibration Failure
For WebMaster
The controller will display a calibration failure if the offset calculated in the
Zero Calibration is outside of the range –20 to 40
mV
or the slope
(mV/ppm) calculated in the One Point Process Calibration is outside of the
range of the nominal mV per 0.1 to 2.0 ppm.
For WDIS:
The acceptable range for the slope (mV/ppm) is the nominal mV per 0.5 to
2.0 ppm. The range of mV for a Zero Calibration is –100 mV to 100 mV.
For W100 or W600 or 900:
The acceptable range for the slope (mV/ppm) is the nominal mV per 0.2 to
10.0 ppm. The range of mV for a Zero Calibration is –100 mV to 100 mV.
To calculate the nominal slope for your sensor, divide the high end of the
nominal range by -2000. For example, for a 0-20 ppm sensor, the nominal
slope is -2000/20 = -100 mV/ppm.
Possible Causes
Corrective Actions
Insufficient conditioning
Wait for between 1 hour and 48 hours before
attempting a calibration.
Insufficient sample flow
Increase flow rate to between 20 and 100 liters
per hour
Air bubbles in sample
Dislodge bubbles
Adjust flow rate higher if necessary
Allow sample to outgas
Loose electrolyte housing
Tighten electrolyte housing
Protection cap not removed
Remove protection cap
Dirty electrodes
Clean electrodes with Emory paper
Replace balls in cleaning attachment
No electrolyte fill solution in
electrolyte housing
Fill electrolyte housing with electrolyte
Air bubbles in electrolyte
Empty electrolyte housing and refill it
Faulty sensor
Replace sensor
Faulty analysis equipment or
reagents
Consult test equipment instructions
Sample contaminated with
interfering molecule (refer to
Sensitivity specification in
section 6.0)
Remove source of contamination
Faulty wiring
Check wiring
Faulty controller sensor input
WebMaster, WDIS4
Go to the Sensor Input menu and perform a
self-test. If this passes, then the problem is with
the sensor or its wiring. If it fails, then
disconnect the sensor from the circuit board and
try the self-test again. If it still fails, replace the
circuit board.
Summary of Contents for WDS100
Page 9: ...6 Typical Installation...
Page 16: ...13 W900...