5.2.1 Contacting
Conductivity
Settings
Touch the Settings icon to view or change the settings related to the sensor.
Alarms
Low-Low, Low, High and High-High Alarms limits may be set.
Deadband
This is the Alarm Deadband. For example, if the High Alarm is 3000, and
the deadband is 10, the alarm will activate at 3001 and deactivate at 2990.
Default Temp
If the temperature signal is lost at any time, then the controller will use the
Default Temp setting for temperature compensation.
Cable Length
The controller automatically compensates for errors in the reading caused
by varying the length of the cable.
Gauge
The cable length compensation depends upon the gauge of wire used to
extend the cable
Cell Constant
Do not change unless instructed by the factory.
Units
Select the units of measure for the conductivity.
Name
The name used to identify the sensor may be changed.
Type
Select the type of sensor to be connected.
5.2.2 Electrodeless
Conductivity
Settings
Touch the Settings icon to view or change the settings related to the sensor.
Alarms
Low-Low, Low, High and High-High Alarms limits may be set.
Deadband
This is the Alarm Deadband. For example, if the High Alarm is 3000, and
the deadband is 10, the alarm will activate at 3000 and deactivate at 2990.
Default Temp
If the temperature signal is lost at any time, then the controller will use the
Default Temp setting for temperature compensation.
Installation Factor
Do not change unless instructed by the factory.
Cable Length
The controller automatically compensates for errors in the reading caused
by varying the length of the cable.
Gauge
The cable length compensation depends upon the gauge of wire used to
extend the cable
Cell Constant
Do not change unless instructed by the factory.
Range
Select the range of conductivity that best matches the conditions the sensor
will see.
Units
Select the units of measure for the conductivity.
Name
The name used to identify the sensor may be changed.
Type
Select the type of sensor to be connected.
Summary of Contents for W600 Series
Page 15: ...12 Figure 1 Conduit Wiring ...
Page 36: ...33 Overview of the menu structure ...
Page 37: ...34 ...
Page 74: ...71 WCT600 Sensor option BD and FD ...
Page 75: ...72 WCT600 Sensor option BN or FN ...
Page 76: ...73 WCT600 Sensor option BA BB BC FA FB FC ...
Page 77: ...74 WCT600 Sensor option DE DF ...
Page 78: ...75 WCT600 Sensor option DN ...
Page 79: ...76 WCT600 Sensor option HA HB and HC ...
Page 80: ...77 WCT600 Sensor option HD ...
Page 81: ...78 WCT600 Sensor option HN ...
Page 82: ...79 WDS600 Sensor option FF or FN ...
Page 83: ...80 WDS600 Sensor option PN ...
Page 84: ...81 WDS600 Sensor option PX ...
Page 85: ...82 WPH600 Sensor option PN ...
Page 86: ...83 WPH600 Sensor option PX ...