Cal Successful or Failed
If successful, touch Confirm to put the new calibration in memory. The calibration adjusts
the offset and the gain (slope) and displays the new values. If failed, you may retry the
calibration or cancel. Refer to Section 8 to troubleshoot a calibration failure.
Resume Control
Replace the sensor in the process and touch Confirm when ready to resume control.
One Point Analog Calibration
OK to disable control?
Touch Confirm to continue or Cancel to abort.
Input Value
Enter the mA value that the transmitter will be sending. Touch Confirm to continue or
Cancel to abort.
Please set input signal to specified value
Make sure that the transmitter is sending the desired mA signal. Touch Confirm to continue
or Cancel to abort.
Automatic circuit calibration in progress
Cal Successful or Failed
If successful, touch Confirm to
save calibration results
. The calculated offset will be
displayed.
If failed, you may retry the calibration or cancel. You may also restore calibration to the
factory defaults. The calibration will fail if the measured mA is more than 2 mA away from
the Input Value entered.
Please restore input signal to process value
Put the transmitter back into normal measurement mode if necessary and touch Confirm
when ready to resume control.
Two Point Analog Calibration
OK to disable control?
Touch Confirm to continue or Cancel to abort.
Input Value
Enter the mA value that the transmitter will be sending. Touch Confirm to continue or
Cancel to abort.
Please set input signal to specified value
Make sure that the transmitter is sending the desired mA signal. Touch Confirm to continue
or Cancel to abort.
Automatic circuit calibration in progress
Second Input Value
Enter the mA value that the transmitter will be sending. Touch Confirm to continue or
Cancel to abort.
Please set input signal to specified value
Make sure that the transmitter is sending the desired mA signal. Touch Confirm to continue
or Cancel to abort.
Automatic circuit calibration in progress
Cal Successful or Failed
If successful, touch Confirm to
save calibration results
. The calculated offset and gain will
be displayed.
If failed, you may retry the calibration or cancel. You may also restore calibration to the
factory defaults. The calibration will fail if the offset is more than 2 mA or the gain is not
between 0.5 and 2.0.
Please restore input signal to process value
Put the transmitter back into normal measurement mode if necessary and touch Confirm
when ready to resume control.
Summary of Contents for W600 Series
Page 15: ...12 Figure 1 Conduit Wiring ...
Page 36: ...33 Overview of the menu structure ...
Page 37: ...34 ...
Page 74: ...71 WCT600 Sensor option BD and FD ...
Page 75: ...72 WCT600 Sensor option BN or FN ...
Page 76: ...73 WCT600 Sensor option BA BB BC FA FB FC ...
Page 77: ...74 WCT600 Sensor option DE DF ...
Page 78: ...75 WCT600 Sensor option DN ...
Page 79: ...76 WCT600 Sensor option HA HB and HC ...
Page 80: ...77 WCT600 Sensor option HD ...
Page 81: ...78 WCT600 Sensor option HN ...
Page 82: ...79 WDS600 Sensor option FF or FN ...
Page 83: ...80 WDS600 Sensor option PN ...
Page 84: ...81 WDS600 Sensor option PX ...
Page 85: ...82 WPH600 Sensor option PN ...
Page 86: ...83 WPH600 Sensor option PX ...