Power-up
B
13
Symbol
Description
Step increment transmission mode
Single ramp signal transmission mode
Cyclic ramp signal transmission mode
Scaling
On hold
Filtering
%FS (full scale) function
Error (over-calibration in measurement or error on
the value emitted…)
Incremental mode using the arrows
Battery life indication
Acquisition in progress (the value on the right of the
pictogram indicates the number of values recorded)
The table below provides a definition of each pictogram of the function keys
Symbol
Description
Tab key
Open a drop-down list
Close a drop-down list
Delete the selected item
Clear the selection
Add the item being edited
B.1.7
Operating modes
There are 2 different operating modes:
Measurement mode (displayed in mV, Ohm, °C or °F)
Simulation mode (value displayed in mV, Ohm, °C or °F).
The configuration of the operating mode is done at the extinction of the product.
At power up, the unit automatically switches measurement mode to avoid possible conflicts if the signal had been
connected to an external circuit.
The functional and electrical characteristics not to be exceeded are described below:
B.1.7.1
Constant voltage measurement
Gauge
+100mV
Resolution (display)
1 µV or 0.01°
Scope of range:
-10 mV to + 100 mV
Scaling
yes
B.1.7.2
Voltage emission
Gauge
+ 80 mV
Resolution (display)
1 µV or 0.01°
Scope of range:
-9.5 mV to +80 mV
Scaling
yes