BMW 2500G
Service tasks
4. To check the zero point of the scales for oil and UV additive, open
the cover (
21
) on the left side and take the containers off the
catches:
– Container for fresh oil (
18
)
– Container for UV additive (
19
) and
– Drained oil receptacle (
20
)
When the scales are relieved, press
ENTER
to confirm. The prompt
to remove the containers from the scale will flash.
✓
Once you have set the zero point, the selection menu for the scales
appears again.
5. Return the containers to the operating positions:
Reattach the containers for oil (
18
) and (
20
) and for the
UV additive (
19
) to the snap locks and close the cover (
21
).
6. Press
STOP
twice to access the standby menu.