79
www.nieaf-smitt.com
Nieaf-Smitt is a brand name of Mors Smitt
7 Active power: Watt (DC/AC)
Function
Range
Accuracy
ACW / DCW
9.999 kW*
99.99 kW
599.9 kW
A,error×V,
V,error×A,reading
* The measured value <1.000 kW, add 10 dgt to the accuracy.
Overload protection
1000 V rms 600 A rms
Accuracy defined for
ACW : Sine wavef , ACV > 10 V rms, AC A
> 5 A rms
Freq. 50~60 Hz, PF=1.00
DCW: DC V > 10 V, DC A > 5 A
8 Power factor
Function
Range
Accuracy
PF
1.00
+ 5 dgt
Overload protection
1000 V rms 600 A rms
9 Resistance & continuity & diode
Fucntion
Range
Accuracy
Resistance
999.9 Ω
9.999 kΩ
99.99 kΩ
± (1.0% + 5dgt)
± (1.0% + 3dgt)
Continuity
999.9 Ω
± (1.0% + 5dgt)
Diode
0.40~ 0.80 V
± 0.1 V
Overload protection
1000 V rms
Max. tests current
± 0.5 mA V
Max. open circuit
voltage for Ω
± 3 V
Max. open circuit
voltage for diode
± 1.8 V
Continuity threshold
<30 Ω beep on / >100 Ω beep off
Continuity indicator
2 kHz tone buzzer
Continuity response
time
< 100 ms