4.6 MATERIALS
NOTICE! Aggressive substances may damage
components, consumables and accessories.
•
Check the chemical resistance before using organic
solvents or aggressive chemicals.
•
Only use liquids whose vapors do not attack the
materials used.
5 PIPETTE OPERATION
5.1 PIPETTING GUIDELINES
4.5 VOLUME INDICATOR DISPLAY EXAMPLE
•
Check the tip cone to make sure it is clean.
•
While using the pipette make sure that operating
plunger is handled slowly and smoothly.
•
Ensure that the tip is firmly attached on the tip cone.
Please check for foreign particles and remove if any,
around the tip cone.
•
Make sure that the temperature of tip, pipette and liquid
are at equilibrium.
•
While aspirating, hold the pipette in upright position and
keep the tips at a constant depth below the surface of
the liquid.
•
Pre-rinse the pipette tip before aspirating the sample by
filling and emptying the pipette tip five (5) times. This is
important when dispensing samples that have a
viscosity and density different from water and for volatile
solvents.
•
Do not pre-rinse the tip when pipetting samples with
temperatures different from the current ambient
temperature. Be sure to change the pipette tip after
each pipetting.
•
For the volatile solvents you should saturate the air
cushion of your pipette by aspirating and dispensing the
solvent repeatedly before aspirating the sample.
•
When pipetting liquids with temperature different to the
ambient temperature, pre-rinse tips tip several times
before use.