6 | Operation
TD22
18 / 27
B-H86.0.0D.DB201-1.0
–
T.OF
= display zero point correction – setpoint zero point
–
T.OF
= -0.2 °C – 0.0 °C = -0.2 °C
–
T.SL
= (setpoint gradient correction / (display gradient correction –
T.OF
) – 1) *100
–
T.SL
= (37.0 °C / (36.6 °C – (-0.2)) -1) *100 = 0.54
Outcome of an action
The changed value is saved and the
Configuration
menu is closed.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.