vwr.com I Instruction manual VWR
®
CO
2
incubator
Possible events:
Date/time change
– data/time were changed
Device On
– the incubator is switched on (on the main switch)
Device Off
– the incubator is switched off (on the main switch)
Door closed
– the door is closed
Door opened
– the door is opened
Open door alarm start
– open door alarm has been activated
Open door alarm stop
– open door alarm has been deactivated
Over Protection Start
– over-temperature protection has been activated
Over Protection Stop
– over-temperature protection has been finished
Program Edit
– changing the program parameters
Program End
– program is finished
Program Restarted
– program has been restarted after power failure
Program Start
– starting the program
Program Stop
– stopping the program
Under Protection Start
– sub-temperature protection has been activated
Under Protection Stop
– sub-temperature protection has been finished
Lower temp. alarm Start
– lower temperature alarm has been activated
Lower temp. alarm End
– lower temperature alarm has been finished
Upper temp. alarm Start
– upper temperature alarm has been activated
Upper temp. alarm End
– upper temperature alarm has been finished
Deleted Measurement
– user measurements have been deleted
Deleted All Measurement
– all measurements have been deleted
User added
– new user has been added
User updated
– user has been changed
User deleted
– user has been deleted
Program saved
– new program has been saved
Program deleted
– program has been deleted
Program updated
– program has been updated
Lower CO2 alarm Start
– lower CO
2
concentration alarm has been activated
Lower CO2 alarm End
– lower CO
2
concentration alarm has been finished
Upper CO2 alarm Start
– upper CO
2
concentration alarm has been activated
Upper CO2 alarm End
– upper CO
2
concentration alarm has been finished
CO2 Pressure low Start –
low CO
2
pressure alarm has been activated
CO2 Pressure low Stop –
low CO
2
pressure alarm has been finished
Sterilization program start –
sterilization program has been started
Sterilization program failed –
error or sterilization program has been stopped / interrupted (incomplete sterilization)
Sterilization program successful –
successful sterilization at 180
o
C for 2 hours has been finished
Sterilization program completed –
sterilization program has been finished (37
o
C in the chamber)
Start program cancelled, door is open
–
sterilization start has been cancelled, the door is open
Info
10.7.
The panel contains the following information:
software versions,
name of the unit,
manufacturer’s address,
manufacturer’s webstie.