5.15.
Defrosting
All models are equipped with a special defrosting program
(Figure 30)
. It appears on the list of programs
available in the main screen after pressing the button
.
Defrosting consists in temporarily turning on the heaters, whose aim is to defrost the coating of ice or frost
on the chamber walls during normal use. Then the chamber is heated to about
30°C and maintained for
30 minutes.
If the defrosting function is active, you will see the following icon
on the main screen.
5.16.
Corrections
In this window
(Figure 31)
you can set temperature correction:
•
The temperature indicated on the display by adding the correction value. The set correction value is
taken in the whole temperature range operation of the device. For example, if the average tempera-
ture displayed by the device indicates 20,0
°C and the average temperature measured by independ-
ent, external sensor indicates 2
0,5°C, the correction should be set on +0,5°C. The average temper-
ature should be calculated from chosen period of time e.g. 30 min. The correction range is between
-
5°C to +5°C.
After defrosting has finished, you need to wipe out the chamber. This will reduce the risk of quick
frosting of the chamber.
The device has been calibrated by the manufacturer in accordance with applicable norms. The
temperature shown on the display corresponds with a great accuracy to the temperature in the
Figure 30
– defrosting program