11
Press
to delete the test result or stored data. If the USB disk was installed, the data will be deleted
from the USB disk, otherwise the data will be deleted in the internal memory.
Blank
Put the reference in the main light path and reference light path, press
to carry out the blank.
Measure samples
Put the sample in the main light path and reference in the reference light path, press
to measure.
Print the test results
Press
to print the test results.
Store the test results
Press
to store the test results, input the file name using the numeric keys and press
to save. If
the USB disk was installed, the data will save in the USB disk, otherwise the data will save in the internal
memory.
Load the stored file
In the test interface, press
to go into the file selecting interface, press
,
to select the file you
want, press
to open. If the USB disk was installed, the data will open from the USB disk, otherwise the
data will open from the internal memory.
Operation
Self-check
Remove all the blocks in the light path and close the lid of the compartment. Switch on the power supply to
begin the self-test.
Warm-up
After self-test, the instrument goes into pre-ready state. For accurate test results, at least 30 minutes of
warm-up is required.
Check the cuvettes