![VWR International 89508-420 User Manual Download Page 21](http://html1.mh-extra.com/html/vwr-international/89508-420/89508-420_user-manual_1057698021.webp)
1. Disconnect incubator for power source.
2. Remove screws from lower back panel, then tilt back panel out.
3. Locate device to be replaced on back panel, then remove mounting
screws.
4. Unplug quick connect terminals, noting positions.
5. Place quick connects on new device and mount on back panel.
6. Replace back panel, then apply power.
7. Refer to General Functional Checks to ensure that new device is
operating.
1. Disconnect incubator from power source.
2. Remove shelves from chamber area.
3. Remove heater cover at top back of chamber by removing 2 screws on
either side.
4. Replace the THM1 or HTR1, if necessary, then coat electrical
connections with RTV102 or other electrical insulator. To replace fan,
remove HTR1, then remove 4 screws that hold enclosure, then replace
fan.
5. Re-assemble in opposite order as previously described.
6. Apply power, refer to General Functional Checks to ensure that new
device is operating.
6-4
Refrigerated Incubator
VWR Collection
Section 6
Maintenance
Replacing SSR1, K1,
K2 or T1
Replacing FAN2,
THM1, HTR1