01 Mode
- Same as 00 Mode, except that the setpoint is locked and
can't be changed.
02 Mode
- Setpoint is locked, plus, the ALARM will activate and the
safety relay will open as follows: chamber < (setpoint -3) or chamber
> (se3)
In order to use this mode of protection, modes 00 or 01 will have to be
used to reach the setpoint temperature. Once the chamber temperature
reaches the setpoint and is stable, protection mode 02 can be set.
This mode can only be used if the door is to remain shut and other
disturbances don't effect the chamber temperature.
As an added safety feature, the incubator has a thermostat in the heater
compartment that limits the upper chamber temperature to 65°C.
Note
If the chamber temperature falls outside the setpoint by ±3°C, the
ALARM will activate and safety relay will interrupt power.
s
Refrigerated Incubator
5-5
VWR Collection
Section 5
Operation
Accessing Controller
Menus (continued)