24 Faults
If the fault cannot be eliminated with the help of the fault table, please inform Customer Service.
Please specify the type of centrifuge and the serial number. Both numbers can be found on the name plate of the
centrifuge.
Perform a MAINS RESET:
Switch off the mains switch (switch position "0").
Wait at least 10 seconds and then switch on the mains switch again (switch position "
").
Message / fault
Cause
Remedy
No display
---
No voltage.
Mains input fuses defective.
Check supply voltage.
Check mains power input fuse,
refer to Chapter "Change mains
input fuse".
Mains switch ON.
1 Faulty
speedometer.
TACHO - ERROR
2
Motor, electronics defective.
CONTROL - ERROR
8
Error in lid locking or lid closure.
Open the cover.
Switch off the mains switch
(switch position "0").
Wait at least 10 seconds.
Turn the rotor vigorously by hand.
Switch on the mains switch again
(switch position "
"). The rotor
must turn during switch-on.
IMBALANCE
---
The rotor is unevenly loaded.
Open lid.
Check the loading of the rotor,
see chapter "Loading the rotor".
Repeat the centrifugation run.
CONTROL - ERROR
4,6
Error in lid locking or lid closure.
N > MAX
5
Rotation too fast
N < MIN
13
Rotation too slow
Perform a MAINS RESET.
MAINS INTERRUPT
---
Power failure during the centrifugation run.
(The centrifugation run was not finished.)
Open lid.
Push
START / PULSE
button.
Repeat the centrifugation run if
necessary.
CONTROL-ERROR 22,
25 - 27
Error / defect electronics
CONTROL-ERROR
23
Error / defect controls
SER I/O - ERROR
30 - 36 Error / defect electronics
° C * - ERROR
51 - 53,
55
Error / defect electronics
FU / CCI - ERROR
60 - 64,
67, 68,
82 - 86
Error / defect electronics / motor
SYNC-ERROR
90
Error / defect electronics
SENSOR-ERROR
91, 92
Error / defect unbalance sensor
KEYBOARD-ERROR
---
Error / defect controls
Perform a MAINS RESET.