SECTION 5 - TROUBLESHOOTING - 3-STAGE
5.6 DIAGNOSTIC TABLES
FURNACE SETUP COMMON ERRORS
Symptom
Possible Setup Error
Display blank
No line voltage
Power switch failure
J8 disconnected
No Setup functions after ENTER
Furnace was in start cycle at power down
Muffle does not heat after power up
Not in start cycle
Door switch not closed
No door movement after switch is pressed
Furnace is in TEST mode
Motor connector J5 loose
POWER SUPPLY VOLTAGES
Output
Voltage
Output
Test
Input
Test
Voltage
Tolerance
Ripple
At
Ripple
At
+12V
+/-350mV
0.02Vac
PT1 & PT4
3Vac
PT1 & PT4
+ 5V*
+/-250mV
0.02Vac
U3-3
3Vac
U3-1
* Note: +5VDC is dependent on the presents of the 12VDC, U3 on the main PC Board regulates the +5VDC.