![VTI SCP1000 User Manual Download Page 10](http://html.mh-extra.com/html/vti/scp1000/scp1000_user-manual_1056070010.webp)
SCP1000 DEMO KIT
User Manual
VTI Technologies Oy
9/19
www.vti.fi Doc.Nr.
8268000.03
Rev.0.03
4.1.5 Pres offset adjustment
In pressure offset adjustment display user
can adjust the absolute pressure offset
value of the SCP1000. The MCU inside the
demo box can read and write the pressure
offset adjustment of the SCP1000 by using
the SCP1000 internal EEPROM registers'
USERDATA1 (0x29) and USERDATA
(0x2A). The offset adjustment is performed
in high resolution measurement mode.
After SCP1000 offset is adjusted in this
display, user can verify the calibration
results in MEAS RESULTS display.
In offset adjustment display user can:
Figure 9. Pressure offset adjustment display.
- Program the SCP1000 offset adjustment in to the SCP1000 internal EEPROM registers
USERADATA1 and USERDATA2 ("Start offset adjustment" button)
-
Read the current SCP1000 offset adjustment data from the two EEPROM registers ("Read
current adjustment data" button)
-
Reset the SCP1000 offset adjustment data by writing zeros in to two EEPROM registers ("Zero
offset adjustment data" button)
Please refer to the following documents:
-
"TN48 SCP1000 Pressure Offset Adjustment" for more detailed information on offset zeroing
and offset adjustment bit level presentation that is used in this demo
- "SCP1000 Product Family Specification 8260800" for more detailed information SCP1000
EEPROM register reading/writing