Ceramic Master E20 en v1
13
7.
IMPORTANT PRACTICAL INFORMATION
1. Avoid positioning of furnace and pump in the
immediate vicinity of heat sources (radiators).
2. Install the vacuum pump in a well-ventilated
locations. Ensure that the apertures in the frame plate
are free and that no foreign object can fall into the
furnace base.
3. Avoid placing any objects on the frame plate, place
only on the face cooling plate.
4. Ensure that the sealing ring in the furnace head and
the sealing rim of the furnace base are kept clean and
undamaged.
5. Close the chamber between working cycles, but not
tightly, and leave a distance of approximately 1 cm.
6. It is strongly recommended to use porous ceramic tray
(on request - № 81022).
WARNING!
For precise operation of the device is
necessary tempering oven for half an hour before the
first use of the day. To do this after turning on the
furnace, run at least one program with an empty camera
and wait another 20 minutes.
7. If the power supply is interrupted during the working
process and:
a.
If the temperature in the chamber has not
decreased by more than 15 degrees, the current
program will continue from the stage at which it
has been stopped.
b.
If the temperature in the chamber has decreased
by more than 15 degrees, the following message
will appear on the display:
LONG TIME POWER OFF
In this case the program cannot continue. You have to
press the button
STOP
in order to exit this state.
The total number of the programs is 203 - with numbers
from 0 to 202. Programs from №1 to №18 have
parameters for ceramic NORITAKE
.
The manufacturer has
entered same parameters in all other programs.
The program №201 is devoted for calibration of the
furnace by means of a silver test.
The program №202 is designed for auto calibration using
the optional calibration kit.
It is responsibility of the customer to enter the data for
the ceramic that he uses before he starts working.