7. S
AFETY
I
NFORMATION
7.1. S
AFETY
F
EATURES
The AC Interface has been designed with several fault-tolerant safety features that provide the operator with a
high level of protection against the risk of electric shock. As with any dangerous equipment, however, it is
important that an assessment of the overall risk to safety is made during installation. It is the user’s
responsibility to ensure compliance with any regulations that may be applicable to the health and safety of
operators.
7.2. S
AFETY
I
NSTRUCTIONS
WARNING:
Voltech’s AC Interface Fixture can pass voltages of up to 600VRMS and 10ARMS during testing.
It is essential that an approved safety interlock system, such as the Voltech light curtain, is installed and
maintained in accordance with the guidelines given in the AT3600/AT5600 user manual.
•
Do not coil any of the high-voltage connection leads.
•
Replace the fuse only with the same type and rating as specified in this manual.
•
Refer servicing only to qualified personnel who understand the danger of shock hazards.
•
Ensure operators are aware of the potential shock hazard.
•
Ensure operators are trained to avoid contacting the part being tested or any exposed parts of the
fixture during testing.
•
Beware of energy stored with the capacitance or inductance of the part being tested when handling
the part after test.
•
Whenever possible use the external source enable line to switch off the external source when not in
use.
•
Ensure that all equipment used is earthed.
•
Ensure that cables are secure.
Maintenance
•
Test the safety interlock system weekly.
•
Examine all test leads for damage to insulation by overheating weekly and replace if required.
•
Return the AC Interface and all leads to Voltech annually for a safety and performance verification.