VOLTCRAFT FG 250D Operating Instructions Manual Download Page 4

Setting the waveform (sine, triangle and square wave signal)

Using the horizontal cross button (4) the waveform can be selected.
Each press of the button „left“ or „right“ switches between waveforms. The current waveform is
shown in the display in the line „Mode“ .
Every change is marked with an asterisk (*) in the display. If the cross button is not used for
longer than 20 seconds the device saves the changes in the internal, non-volatile memory.

Stepwise setting the output frequency

Using the vertical cross button (4) the output frequency can be selected.
Every time the buttons „up“ or „down“ are pressed, the frequency is stepped by 1 Hz.
If the button is kept pressed for a longer time, a quick adjustment is performed. The current 
frequency is shown in the display in the line „FRQ=“

Directly setting the output frequency

The frequency can also be entered directly for every decimal place. To do so, press the centre
of the cross button. A setup menu opens. Select the appropriate decimal using the cross 
button (left/right); it is marked by a flashing symbol. 
Change the value of the decimal using the cross button (up/down). After setting the correct 
frequency, confirm your input by pressing the centre of the cross button („Enter“). The set menu
closes.

Setting the amplitude

The output voltage for the sine or the triangle wave can be set using the rotary knob „AMPLI-
TUDE“ (5).
For the waveform „square/TTL“ this rotary knob is not needed. Set it to middle position for the
waveform TTL.

Setting the offset voltage/pulse width

For the sine and the triangle wave the offset adjustment is done using the „OFFSET“ control
(6). 
When using „TTL signal (square wave)“ this control has a different function. In this case, the
„OFFSET“ control is used to set the pulse width for the TTL signal.

Connecting the measuring leads

Only use measuring accessories suited to the specifications of the generator
(e.g. impedance 50 Ohm). 
The BNC output is voltage-free. Do not apply and external voltage to the BNC
socket.
Never short-circuit the signal output.
Signal lines must be shielded and must not exceed a total length of 3 metres.
Ensure a good earth contact.

Connect the BNC measurement lead to the BNC output socket „OUTPUT“ (7) and lock the 
bayonet connector.

Contact the circuit with the other end of the measurement lead.

Maintenance and Cleaning

Apart from occasional cleaning this device is maintenance-free. 
Use a dry, clean, lint-free, antistatic cloth to clean the device. Do not use any abrasive or chem-
ical agents or detergents containing solvents.

Always unplug the mains plug from the mains socket before cleaning the unit.

Troubleshooting

By purchasing waveform generator you have acquired a product with state of the art design that
is operationally reliable. Nevertheless, problems or faults may occur. Which is why we would
like to describe here how to solve any problems:

Always follow the safety instructions!

Problem

Solution

The operation indicator 

No main’s voltage? 

does not flash.

• Check the safety switch for the mains socket
• Check whether the mains plug is seated correctly in the

mains socket

• Is the power switch locked in position „I“

No or only very 

Is the output overloaded (too low impedance)?

low output signal

Is the amplitude set too low? Increase the signal voltage using
the rotary knob „AMPLITUDE“.

The output signal 

The measuring circuit has a higher impedance.

is slightly distorted 

If necessary, use a 50 Ohm resistance

output.

in order to adjust the circuit impedance. 

Repairs, other than those just described, should only be carried out by an
authorised specialist. Check the technical safety of the device on a regular
basis e.g. for damage to the housing etc. Any changes and repairs you make
on the device will render the guarantee/warranty null and void.

Disposal

Electronic products are raw material and do not belong in the household waste.
When the device has reached the end of its service life, please dispose of it,
according to the current statutory requirements, at your local collecting site.
Disposal in the domestic waste is not permitted!

Technical Data

Output signals ......................................Sine, square wave (TTL), triangular

Frequency range ..................................1Hz - 250 kHz

Step ......................................................1Hz

Accuracy ..............................................0.01% ± 0.1Hz

Amplitude (sine, triangle)......................100 mVss - 5 Vss

Offset (sine, triangle) ............................-2.5 V to +2.5 V

Distortion ratio (sine) ............................< 0.1 % (1000Hz), <0.5 % (20kHz)

Rise/decay time (TTL) ..........................<10 nS

Output impedance ................................50 Ohm

Operating voltage ................................230 V/AC (+/- 10 %) 50 Hz;

Max. power ..........................................3 W

Operating temperature ........................5 to 40 °C

Storage temperature ............................-20 to +70 °C

Dimensions (W x H x D) ......................225 x 73 x 212 mm

Weight ..................................................0.95 kg

Nominal data with tolerances apply after a warm up time of at least 30 min-
utes at an ambient temperature of 23 °C. Values without tolerances are guid-
ance values for an average device.

Impressum /legal notice in our operating instructions

These operating instructions are a publication by Voltcraft

®

, Lindenweg 15, D-92242 Hirschau/Germany, Phone +49 180/586 582 7 (www.voltcraft.de). 

All rights including translation reserved. Reproduction by any method, e.g. photocopy, microfilming, or the capture in electronic data processing
systems require the prior written approval by the editor. Reprinting, also in part, is prohibited. 

These operating instructions represent the technical status at the time of printing. Changes in technology and equipment reserved.

© Copyright 2010 by Voltcraft

®

Summary of Contents for FG 250D

Page 1: ...nn dies von Hand m glich ist k nnen spannungsf hrende Teile freigelegt werden Kondensatoren im Ger t k nnen noch geladen sein selbst wenn das Ger t von allen Spannungsquellen getrennt wurde Beachten S...

Page 2: ...e chemische und l sungsmittelhaltige Reinigungsmittel Ziehen Sie immer den Netzstecker aus der Netzsteckdose bevor Sie das Ger t reinigen Behebung von St rungen Mit dem Funktionsgenerator haben Sie ei...

Page 3: ...If you open covers or remove parts unless you can do this by hand life parts may be exposed Capacitors inside the device may still be charged even if the device has been disconnected from all voltage...

Page 4: ...lug from the mains socket before cleaning the unit Troubleshooting By purchasing waveform generator you have acquired a product with state of the art design that is operationally reliable Nevertheless...

Page 5: ...e que des parties sous tension ne soient plus prot g es sauf si l on peut les ouvrir manuellement Les condensateurs de l appareil peuvent encore tre charg s m me lorsque l appareil a t d connect de to...

Page 6: ...nnage Avec le g n rateur de fonction vous avez acquis un produit la pointe du d veloppement tech nique actuel et b n ficiant d une grande s curit de fonctionnement En revanche il est pos sible que des...

Page 7: ...onderdelen behalve als dit handmatig mogelijk is kunnen onderdelen onder spanning worden blootgelegd Condensators in het apparaat kunnen nog geladen zijn ook als het apparaat van alle spanningsbronne...

Page 8: ...stekker altijd uit de contactdoos voordat u het apparaat schoon maakt Verhelpen van storingen U hebt met de functiegenerator een product aangeschaft dat volgens de nieuwste stand der techniek is ontwi...

Reviews: