62
Measurement accuracy
Statement of accuracy in ± (% of r display error in counts (= number of smallest
points)). The accuracy is valid for one year at a temperature of +23 °C (± 5 °C) and at a rela-
tive humidity of 45% - 75% RF, non-condensing.
Temperature coefficient (outside this temperature range): +(0.1 x accuracy) / °C.
Voltage measurement
Function
Range
Accuracy
Addendum
DC-V
600.0 V
±(2% +3)
AC-V
600.0 V
±(2% +3)
50 - 400 Hz
Overload protection: 600 Vrms
Transducer AC-V: AC-coupled, actual effective value (True RMS)
Input impedance: 10 MΩ / <100 pF
Crest factor for non-sinusoidal voltages:
1.0 - 2.0
+3% to the base accuracy
2.0 - 2.5
+5% to the base accuracy
2.5 - 3.0
+7% to the base accuracy