11. RS232 port
The meter is fitted with a serial interface for exchanging data with a computer. This can be found
on the right hand side under the lid. The interface is in the form of a 3.5mm jack plug socket and
requires a compatible data cable, which is an optional extra.
The data cable has the following components:
Jack plug 3.5 mm mono
9-pole Sub D socket for PC (serial port)
Centre contact →
Pin 4
External contact →
Pin 2
A 2.2 KOhm is required between pin 2 and pin 5.
The serial data signal is made up of 16 Bits in the following order:
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
Each data bit has the following function:
D15
Start symbol
D14
4
D13
When the upper display is output = 1
When the lower display is output = 2
D12+D11
Displayed unit of measurement: mg/L = 07 // %O2 = 06 // °C = 01 // °F = 02
D10
Polarity; 0=positive; 1=negative
D9
Decimal point (DP) at the appropriate place (right to left); 0 = no DP; 1 =
1DP; 2 = 2DP; 3 = 3DP
D8 to D1
Measured value (D8 = largest digit (MSD), D1 = lowest digit (LSD)). If the
display shows 1234, the following bit set (D8 – D1) will be used: “00001234”
D0
End symbol
The RS232 data format is:
9600, N, 8, 1
Baud rate:
9600
Parity bit:
No parity bit (N)
Data bit number:
8
Stop bit:
1 stop bit