Direct voltage V/DC
Range
Resolution
Accuracy
600.0 mV*
0.1 mV
±(0.8% + 3)
6.000 V
0.001 V
±(0.6% + 3)
6.000 V AC+DC
0.001 V
±(2.0% + 5)
60.00 V
0.01 V
±(0.8% + 3)
60.00 V AC+DC
0.01 V
±(2.0% + 5)
600.0 V
0.1 V
±(0.8% + 3)
1000 V
1 V
*only available via the measuring function "mV"
Overload protection 1000 V; Impedance: 10 MΩ (mV: ≥1000 MΩ)
When the measuring input is short-circuited in the mV area, a display of 5 counts is possible;
For AC+DC function, the frequency range for ACV is 40-400 Hz
Alternating voltage V/AC
Range
Resolution
Accuracy
600.0 mV*
0.1 mV
±(1.3% + 5)
6.000 V
0.001 V
±(1.0% + 4)
60.00 V
0.01 V
±(1.3% + 4)
600.0 V
0.1 V
750 V
1 V
600.0 V "LoZ"
0.1 V
±(2.6% + 3)
600.0 V "Motor"
0.1 V
±(2.0% + 7)
6.000 V - 750 V "LPF"
depending on measurement range
±(2.0% + 3)**
*only available via the measuring function "mV"
**Additional measuring error with the "LPF" function activated
Frequency response 40 Hz - 5 kHz (when the measured voltage frequency is greater than 1 kHz, the accuracy
needs to add 5 Counts); Overload protection 750 V; Impedance: 10 MΩ (mV: ≥10 MΩ)
Specified measurement range: 5 - 100 % of the measured range
When the measuring input is short-circuited, a display of 5 counts is possible
After use of the LoZ function, a regeneration time of 1 minute is required
TrueRMS crest factor (CF) ≤3 CF to 600 V, ≤1.5 CF to 750 V
TrueRMS crest factor for non-sine-shaped signals plus tolerance addition:
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%
Criteria for phase recognition in the "Motor" mode: >80 V/AC, 40 - 80 Hz