68
Frequency “Hz” (electronic)
Range
Resolution
Accuracy
60 Hz - 6.000 kHz*
0.001 kHz
±(0.1% + 6)
60.00 kHz
0.01 kHz
600.0 kHz
0.1 kHz
6.000 MHz
0.001 MHz
40.00 MHz
0.001 MHz
Signal level (without direct voltage share):
≤100 kHz: 200 mV - 30 Vrms
>100 kHz - <1 MHz: 600 mV - 30 Vrms
≥1 MHz - <10 MHz: 1 V - 30 Vrms
10 MHz -40 MHz: 1.8 V - 30 Vrms
*The frequency measuring range starts at 60 Hz
Frequency “Hz” (electrical, subfunction of A and V)
Range
Resolution
Accuracy
40 - 5 kHz
0.1 Hz
Not specified
Signal level: Sensitivity at least 30 Vrms
Diode test
Test voltage
Resolution
approx. 3.0 V/DC
0.001 V
Overload protection: 1000 V; test current: 0.8 mA typ.
Acoustic continuity tester
Measurement range
Resolution
100 Ω
0.1 Ω
≤10 Ω permanent sound; ≥50 Ω, the buzzer won´t beep
Overload protection: 1000 V
Test voltage approx. -3.2 V
Test current -0.7 mA
Do not exceed the maximum permitted input values. Do not touch any circuits or parts of circuits if
they may be subject to voltages higher than 33 V/ACrms or 70 V/DC! Danger to life!