Page: 10 / 28
© 2016 by Vogel & Plötscher GmbH & Co. KG | RCAD | Version 2.0 | 25.01.2015
3.3.4 Initialization of inclination sensor
Procedure
Put the device on the track and switch on the device. Press and hold button 1 until the dis-
play will show a rotating square (
à
Fig. 4a).
Wait about 10 seconds then press button 1 again and wait for the rotating square is chang-
ing its running direction (
à
Fig. 4b). Now lift up the device, turn it by 180° and place it on
same track position again. Wait for about 10 seconds before pressing button 1 again. The
new sensor setting will be saved while the current cant value is shown in the display.
Now perform the roll-over test (
à
Section 3.3.3) again to check if the situation has been
improved.
In case the initialization procedure fails to deliver any improvement, the measuring device
must be checked and recalibrated at the factory.