![VOCARE vitals360 User Manual Download Page 114](http://html.mh-extra.com/html/vocare/vitals360/vitals360_user-manual_1043425114.webp)
Instructions for Use
Page
114
of
117
VITALS360 ® device
is used exceeds the applicable RF compliance level above, this
VITALS360 ® device
should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating theis
VITALS360 ® device
.
c
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between
portable and mobile RF communications equipment and this VITALS360 ® device
This
VITALS360 ® device
is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of this
VITALS360 ® device
can help
prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and this
VITALS360 ® device
as recommended
below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitterm
Rated
maximum
output of
transmitter
W
150 kHz to 80 MHz
outside ISM and
amateur radio bands
150 kHz to 80 MHz
in ISM and amateur
radio bands
80 MHz to 800 MHz
800 MHz to 2.7
GHz
0.01
0.12
0.20
0.035
0.07
0.1
0.38
0.63
0.11
0.22
1
1.2
2.00
0.35
0.70
10
3.8
6.32
1.10
2.21
100
12
20.00
35
70
For transmitters rated at a maximum output power not listed above the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
P
V
d
]
5
.
3
[
1
=
P
V
d
]
12
[
2
=
P
E
d
]
5
.
3
[
1
=
P
E
d
]
7
[
1
=
Summary of Contents for vitals360
Page 1: ...Instructions for Use Page 1 of 117 ...
Page 20: ...Instructions for Use Page 20 of 117 02 DEVICE OVERVIEW 02 1 Know Your Monitor a Device ...
Page 36: ...Instructions for Use Page 36 of 117 Home Button Back Button returns to the previous screen ...
Page 37: ...Instructions for Use Page 37 of 117 03 6 Symbols Bank Home Page LANGUAGES ...
Page 56: ...Instructions for Use Page 56 of 117 04 9 Troubleshooting ...
Page 61: ...Instructions for Use Page 61 of 117 05 3 Display Symbols Bank Pulse Rate Waveform ...
Page 72: ...Instructions for Use Page 72 of 117 ...