Important Control Solution Information
Use only VocalPoint control solutions.
▪
Check the expiration date on the control solution vial.
▪
Do not use if expired.
Control solution, meter, and test strips should come
▪
to room temperature (68 to 77°F/20 to 25°C) before
testing.
Shake the vial, discard the first drop of control
▪
solution, and wipe off the dispenser tip to ensure a
pure sample and an accurate result.
Use only for 90 days after first opening. Record the
▪
discard date (date opened plus 90 days) on the
control solution vial. Discard after 90 days.
Store the control solution tightly closed at
▪
temperatures from 36 to 86°F (2 to 30°C). Do not
freeze.
PLEASE NOTE
The control solution range printed on the test strip vial is
for
VocalPoint
control solution only. It is used to test meter
and test strip performance. It is not a recommended range
for your blood.