56
▼
Operational Problem
1. If the meter does not display a message after
inserting a test strip:
2. If the test does not start after applying the sample:
Battery exhausted.
Replace the batteries.
Battery incorrectly installed or
absent.
Check that the battery is
correctly installed.
Test strip inserted upside down
or incompletely.
Insert the test strip correctly
with the contact bar end first
and facing up.
Defective meter.
Please contact the place of
purchase for assistance.
Insufficient blood sample.
Repeat the test using a new
test strip with larger volume of
blood sample.
Defective test strip.
Repeat the test with a new test
strip.
Sample applied after automatic
shutdown (3 minutes after last
user action).
Repeat the test with a new test
strip.Apply sample only when
flashing
“ ”
appears on the
display.
Defective meter.
Please contact place of
purchase for assistance.
Summary of Contents for SMS-4250
Page 1: ...SMS 4250 Blood Glucose Monitoring System Version 1 0 June 2009 311 4250200 002 Owner s Manual ...
Page 62: ...62 NOTES ...
Page 63: ...63 NOTES ...