WHISPER-0

USER MANUAL

www.vicidbs.com

Version 1.03.0000

6. OPERATION

6.1. Starting the Unit

1. 

Connect the Air Inlet “B” (Fig-1) to the Air line from the compressor (oil free)

2. 

Install a piece of tubing from the Nitrogen Outlet “E” to the LC-MS

3. 

Slowly open the red tap upstream from the coalescent filter “B”

4.  The generator is ready to be used

6.2. Stopping the Unit

To stop the generator, just close the red tap upstream from the coalescent filter “B”

7. TROUBLESHOOTING

ERROR

POSSIBLE CAUSE

POSSIBLE SOLUTION

Nitrogen flow insufficient or lacking

Internal filters are contaminated

Replace the filters

The Whisper-0 is off

Start the Whisper-0

There is a leak somewhere

Check for possible leaks

Nitrogen outlet obstructed

Check and clear the nitrogen 

outlet

Oxygen concentration is too high

Room temperature is less than  

10º C

Raise the room temperature

Membrane no longer functions 

correctly

Membrane column must be 

replaced

5.4. Connecting the Whisper-0 to the LC-MS

ATTENTION:

 Check that the tubing between the Whisper-0 and the LC-MS is free from dust, grease, drops of 

liquid etc, before making the connection. 
Connect the Whisper-0 and the LC-MS with stainless steel tubes.

Summary of Contents for DBS WHISPER 0-120

Page 1: ...USA Canada Central South America techsupportusa vicidbs com Italy service vicidbs com VICI AG INTERNATIONAL Parkstrasse 2 CH 6214 Schenkon Switzerland tel 41 41 925 6200 fax 41 41 925 6201 web www vi...

Page 2: ...NG 4 4 SPECIFICATIONS 5 4 1 TECHNICAL SPECIFICATIONS 5 4 2 SPARE PARTS 6 4 3 SEPARATION PRINCIPLE 6 4 4 COMPONENTS 7 4 5 PROCESS DIAGRAM 8 5 INSTALLATION 8 5 1 TRANSPORT 8 5 2 LOCATION 8 5 3 CONNECTIN...

Page 3: ...including drawings tables and technical descriptions remain the property of the manufacturer and may not be used with the exception of the purposes herein copied reproduced or published without explic...

Page 4: ...perator action in order to avoid undesirable consequences The manual must be consulted in all cases where this symbol is marked on the device Risk of asphyxia Caution is necessary when operating the d...

Page 5: ...ety Instructions The Whisper 0 ejects oxygen as waste Oxygen concentration in the air could change around the generator in a small or not sufficiently ventilated room If fuel gases are present their e...

Page 6: ...ressor air inlet 240 420 490 Pressure drop barg psig 1 5 22 Inlet air quality Clean dry compressed air ISO8573 1 2010 Class 1 2 1 Technology Membrane Warm up time minutes None Electrical supply None N...

Page 7: ...rs DB WN2 002 Pneumatic Valve DB WN2 003 Manometer DB WN2 005 10 liter tank DB WN2 006 Tube 10 m DB WN2 014 Valveless N2 small membrane DB WN2 015 Filter housing for A0035 element without fittings DB...

Page 8: ...en has a slow permeation speed and passes slowly through the membrane wall The resulting nitrogen flow C is then forced out of the membrane and used directly 4 4 Components A D B C E A Active Carbon F...

Page 9: ...unlight Keep away from heat sources Room with adequate aeration ventilation Connect a tube for waste humidity from filter B 5 3 Connecting the Whisper 0 to the Air Compressor There are 2 configuration...

Page 10: ...POSSIBLE SOLUTION Nitrogen flow insufficient or lacking Internal filters are contaminated Replace the filters The Whisper 0 is off Start the Whisper 0 There is a leak somewhere Check for possible lea...

Page 11: ...ter B 3 Let the generator depressurize till the pressure is 0 bar 4 Remove the cap unscrew it from the main body 1 and pull it down 2 5 Remove the filter cartridge by pulling it down 3 6 Replace the f...

Page 12: ...MENT AND THE INFORMATION WHICH IT CONTAINS SHALL NOT BE USED EXPLOITED OR SOLD AND SHALL NOT BE REVEALED OR DISCLOSED TO OTHERS WITHOUT THE EXPRESS WRITTEN PERMISSION OF VICI THIS DOCUMENT SHALL REMAI...

Reviews: