Electromagnetic Compatibility (EMC) Tables
Table 2
This test level assumes a minimum distance between the Continuous ECG Recorder and power source with magnetic field of at
least 15 cm. If the RISK ANALYSIS shows that the Continuous ECG Recorder will be used closer than 15 cm to power source with
magnetic field, the IMMUNITY TEST LEVEL shall be adjusted as appropriate for the minimum expected distance.
Immunity Test Levels
Phenomenon and
Basic EMC standard
or test method
Electrostatic discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV , ± 4 kV, ±8 kV, 15 kV air
30 A/m
50Hz or 60Hz
Radiated RF EM fields
IEC 61000-4-3
Rated power frequency
magnetic fields
IEC 61000-4-8
3 V/m
80 MHz - 2.7 GHz
8% AM at 1kHz
10 V/m
80 MHz - 2.7 GHz
8% AM at 1kHz
Professional Healthcare Facility Environment
(E.g. physician offices, dental offices, clinics etc.)
Home Healthcare Environment
(E.g. restaurants, cafes, shops, etc.)
c)
c)
15