Application Examples
75
In many applications there is only a requirement for one set of measurement results, and
these are not usually very extensive in size. In these cases, the user may program a single
Transmit Buffer Bank to provide the required information. It is recommended that the
user choose the default bank, Bank #0, in this situation.
Each bank has a 6000-character maximum data size and a 50-definition limit. Use
multiple banks if –
•
The amount of data required from the 2553/ 2553E is extensive,
•
The user requires different data dependant on the stage of the application,
•
The user requires different data dependant on other events, which may occur.
In the rare situation where more than 5 banks are required then these will have to be
programmed during the main execution of the application, rather than being programmed
during the initialization as shown here.
After deciding on the data which will be required from the 2553/ 2553E, and possibly
which data will be required from which bank, then the user should program the respective
banks using the BANK0, BANK1, BANK2, BANK3, and BANK4 commands. To
ensure that an unused bank does not contain a definition which would take the 2553/
2553E time to format, it is recommended that unused banks be cleared by sending the
respective BANKn command with no = character and no subsequent data. For
readability of code, and to ensure that the 512-character command limit of the 2553/
2553E is not exceeded, it is recommended that each BANKn command be sent
individually, i.e., each terminated by the <NL> character. Some examples of bank
measurement result definitions are given below.
SINGLE BANK USAGE
Sending the command:
BANK0=VOLTS[A/RMS]/VOLTS[A/CF]/VOLTS[A/THD]/WATTS[A/RMS]<NL>
Configures the 2553/ 2553E to return the RMS Voltage, Voltage Crest Factor, Voltage
Total Harmonic Distortion and Total Load Power respectively (all for phase A) when
bank #0 is read. In this application none of the other banks will be required, thus the
following command will disable them.
BANK1;BANK2;BANK3;BANK4<NL>