Indication of emergency situations during the generator IFL-1210G operation
Indication
Reason
Implication
Er 10
Signal reached minimum
Incorrect actions of the operators
when the level of output signal is
changed manually with the buttons
« » or « ». The transmission is
not stopped
Er 11
Signal reached maximum
Er 12
Power reached maximum
Er 14
Current in load reached maximum
Er 20
Encountered external power voltage not
allowed for charging
Indication of emergency situations
resulting in automatic switching the
transmission off
Er 21
External power voltage during generation
has been too low
Er 22
Voltage of one of internal accumulators
has been too low
Er 23
Voltage of external power has been too
high
Er 30
Current mode did not match to
“presence/absence” of transmitting
antenna due to incorrect connection
Er 40
Highest allowed current in output cascade
has been exceeded
Er 41
Highest allowed consumption current has
been exceeded