vita vPad comfort Operating Manual Download Page 17

17

3. Operation

 

 

4. Materials and firing programs 

5. Firing window 

3. Operation

   

4. Materials and firing programs 

5. Firing window 

Note: 

• 

What? Here you will find all the pre-installed materials.

• 

What for? Saves time, thanks to pre-installed materials.

4. Materials and firing programs 

Working 
instructions 

Manage materials

Note:

• 

What? Here you will find all the working instructions for the VITA materials integrated into the software.

• 

What for? Quick assistance, thanks to pre-installed working instructions.

4.1 Working Instructions 

2

1

1

Select "Working Instructions."

Select "Materials/firing programs."

ADDITIONAL

  

INFORMA

TION

SCOPE OF SUPPL

Y/

VIT

A SYSTEM SOLUTIONS

TECHNICAL DA

TA

SETTINGS

FIRING WINDOW

MA

TERIALS AND

  

FIRING PROGRAMS

OPERA

TION

FIRST USE

THE SYSTEM

Summary of Contents for vPad comfort

Page 1: ...VITA vPad comfort Operating manual VITA perfect match Date of issue 09 18 VITA shade determination VITA shade communication VITA shade reproduction VITA shade control...

Page 2: ...eed sintering furnace The VITA vPad comfort is a state of the art control unit that complies with all international safety standards In order to ensure that you can operate the VITA vPad comfort safel...

Page 3: ...1 Firing programs for single and two stage curves 27 5 2 Firing program Pressing curve 33 5 3 Firing program Sintering curve 36 5 4 Changes after program startup 39 6 Settings 6 1 Setting up the dent...

Page 4: ...4...

Page 5: ...A control unit for controlling a firing device The VITA vPad comfort is not compatible with the VITA SWITCHBOX or the VITA MultiPump ADDITIONAL INFORMATION SCOPE OF SUPPLY VITA SYSTEM SOLUTIONS TECHNI...

Page 6: ...6 1 The system 2 First use 1 1 The control unit 1 USB input Port for connecting cable Adjustable stand 10 inch TFT display...

Page 7: ...nection Vacuum tube connection Pressure regulator 3 bar Connection for VITA SWITCHBOX Connection for VITA SWITCHBOX Compressed air connection max 5 bar Status display Status display 2 2 Connections on...

Page 8: ...vacuum pump VITA VACUMAT 6000 M VITA VACUMAT 6000 MP only 3 Connect the vacuum hose for the VITA vacuum pump VITA VACUMAT 6000 M VITA VACUMAT 6000 MP only 4 Connect the mains power cable 5 Switch on t...

Page 9: ...to enlarge or contract the image Scrolling Swipe downwards or upwards to move between pages in a PDF Short tap You open the basic firing window e g for a firing program with a short tap Rotating Phot...

Page 10: ...als and firing programs Materials firing programs Altering the lift position Settings Profiles Photoviewer Help Active profile Firing device status Switching off the system VITA Energy Efficiency Comb...

Page 11: ...creases the firing chamber temperature to the set standby temperature 500 C by default To deactivate the standby temperature short tap the button again Once the standby temperature has been reached it...

Page 12: ...he VITA profile cannot be modified You can adjust the Lab profile to suit your individual needs There is no option for adding new profiles What for The VITA profile allows you to use the VITA programs...

Page 13: ...e to images in the vPad comfort What for Allows you to customize your screensaver and wallpaper 3 8 Photoviewer Example Importing images into the Screensaver folder 3 6 2 5 1 4 7 Press and hold to sel...

Page 14: ...rt Select Delete Press and hold your finger down on the first image until it is selected Short tap on other images to select them as well Press and hold on the first image until it is selected Short t...

Page 15: ...e Pressing the Help button in the Material overview or the Settings opens the Help function with a list of VITA operating manuals In the Material overview Help shows all the working instructions for V...

Page 16: ...16 3 Operation 4 Materials and firing programs 5 Firing window...

Page 17: ...terials Note What Here you will find all the working instructions for the VITA materials integrated into the software What for Quick assistance thanks to pre installed working instructions 4 1 Working...

Page 18: ...s not possible to make changes to the VITA profile The Lab profile is a copy of the VITA profile and can be modified to suit your needs What for Allows you to design custom materials and material sequ...

Page 19: ...curve types Pre drying curve e g drying stained zirconium restorations Two stage curve special programs e g IPS e max CAD crystallization Sintering curve Pressing curve Single stage curve standard fi...

Page 20: ...ing material 3 3 3 2 2 2 1 1 1 4 Confirm The material can be edited Your final copied material Select Delete material Select the material you want to edit then Edit material Select Copy material Selec...

Page 21: ...t be displayed in the corresponding file name When it is imported onto a vPad the correct material name will be displayed again Example Drag drop 2 1 Your final moved material Press and hold your fing...

Page 22: ...ort a material a USB drive containing a file in 2vtm format must be connected to your vPad comfort Example Importing material 3 6 2 5 1 4 Select the material you want to import The TEST material has n...

Page 23: ...for Allows you to design custom firing programs and firing program sequences 4 3 Managing firing programs Back Create new firing program 3 2 1 The firing program management menu Select Manage firing...

Page 24: ...5 1 1 1 4 Enter a firing program name Your final renamed firing program Select Create firing program Enter a program name and confirm Your final copied firing program Your final firing program Select...

Page 25: ...rm Your final moved firing program Press and hold your finger on the firing program you want to move until it turns transparent and can be moved Select a firing program then click on Delete firing pro...

Page 26: ...l not be displayed in the corresponding file name When it is imported onto a vPad the correct program name will be displayed again Once the export is complete the firing programs in question will be o...

Page 27: ...device back on switch the main switch off and then on again Night mode must be activated before starting a firing program The two stage firing curve has an extra gradient rise time and firing tempera...

Page 28: ...can subsequently be exported onto a USB drive and then transferred to the FDS Firing Data System the VITA program used to analyze the firing data on the PC The firing data system allows you to displa...

Page 29: ...Value for c is determined automatically Modifiable parameters Pre vacuum time 00 00 min 60 00 min Vacuum value 0 100 Please note The time values for the individual lift positions depend on the total...

Page 30: ...layed above the number field The same procedure for entering and altering values applies for the program values in the second rise time on two stage firing curves For reliable firing results with a go...

Page 31: ...ressed The actual values to enter are displayed above the number field Deactivate cooling Press Activate cooling button For slow cooling of the restoration to prevent thermal stress on the material Fi...

Page 32: ...re displayed above the number field This function is available when the cooling temperature is active It adjusts the speed at which the cooling temperature is reached The smaller the furnace opening t...

Page 33: ...The main vacuum cannot be altered in the pressing program The vacuum is controlled automatically throughout the entire pressing time Night mode Back Save Program stop Program start Rise time gradient...

Page 34: ...essing time and pressing pressure 1 2 1 1 2 Note Pressing time 0 00 min 120 00 min Pressing pressure The pressing pressure is changed using the Change pressing pressure button The indicated pressing t...

Page 35: ...ressing object Note Pressing time 0 00 min 120 00 min Pressing pressure The pressing pressure is changed using the Change pressing pressure button Pressing time 2 runs in accordance with the set time...

Page 36: ...service life should be expected The possible input values for the individual phases are displayed above the number field Night mode Back Save program parameters Stop firing program Start firing progr...

Page 37: ...firm Activate first phase Once you have done this confirm Select firing temperature holding time Select first phase Gradient rise time Note To deactivate pre dry and the first phase remove the corresp...

Page 38: ...3 remove the corresponding checkmarks and confirm Save your change Activating deactivating controlled cooling 3 6 2 5 1 4 7 8 9 Confirm Confirm Select first cooling phase Controlled cooling Select se...

Page 39: ...allowed such as changing the rise time which causes an automatic adjustment of the vacuum time 5 4 Changes after program start Please note You can only change inactive program areas The program areas...

Page 40: ...40 4 Materials and firing programs 5 Firing window 6 Settings...

Page 41: ...gs performing updates and viewing operating data 6 Settings Device data Update Cancel settings Operating manual help Screensaver settings Date and time settings Dental furnace settings Pressing pressu...

Page 42: ...ut a minus or plus sign before it automatic plus The correction value corresponds to about 1 C and is entered in the input field for temperature adjustment offset VITA ZYRCOMAT 6000 MS temperature cal...

Page 43: ...e settings ensure that the device is automatically vented at regular intervals This relieves residual pressure in the system and ensures that the pressure readings are correct You can carry out manual...

Page 44: ...roduced an unsatisfactory firing result for example or an error has occurred in the program sequence the operating data the device information or even all the device data can be exported to a USB driv...

Page 45: ...this tap the Change password button and confirm twice without entering a new password 6 6 Setting the screensaver 2 1 Select Screensaver settings Note This setting allows you to activate a screensaver...

Page 46: ...t Language settings Select Date settings Note Here you can adjust the language temperature unit and time and date format Note It is important to enter the correct date and time in order for the firing...

Page 47: ...and date and time format Doing this will delete all your own program settings and custom profiles Note You can set whether you want the confirmation query Do you want to stop the firing process to ap...

Page 48: ...48 5 Firing window 6 Settings 7 Technical data...

Page 49: ...nel when using equipment modified by the user are not permitted Authorized modes of operation The operation of the device is only permitted if this operating manual has been completely read and unders...

Page 50: ...n though the display will remain out of service for the duration of the power failure If the power failure lasts longer than 15 seconds on the VITA VACUMAT 6000 M and MP or longer than 2 minutes on th...

Page 51: ...VITA Easyshade V and for traditional shade determination the VITA Linearguide 3D MASTER or VITA classical A1 D4 can be used For CAD CAM fabrication VITA offers you highly esthetic feldspar and glass...

Page 52: ...be triggered by sev eral causes Recognizing correcting these errors requires specialized knowledge and technical measur ing equipment Replace CPU circuit board Temperature sensor is defective Program...

Page 53: ...r Temperature sensor defective Running program is stopped heating off Inspect connections to the circuit board Inspect replace ventilator Relay error Running program is stopped heating off Replace pow...

Page 54: ...S Empress CAD Celtra Ceram Celtra Press and n ce However it should be noted that the firing processes for IPS e max CAD must be carried out on an Ivoclar Vivadent firing tray The firings for IPS Empre...

Page 55: ...ions you can contact our technical specialists Mr Ralf Mehlin and Mr Daniel Schneider Phone 49 0 7761 56 22 22 Fax 49 0 7761 56 24 46 8 am to 5 pm CET e mail info vita zahnfabrik com VITA instruments...

Page 56: ...A Modulbox is not necessarily a component of the product Date of issue of this information 09 18 After the publication of this information for use any previous versions become obsolete The current ver...