EMC COMPATIBILITY
Electromagnetic compatibility (EMC) is assessed with reference to the following standards:
Guidance and
- electromagnetic emissions - for all ME
EQUIPMENT and ME SYSTEM.
electromagnetic emissions
The ENDOULTRA is intended for use in the electromagnetic environment specified below. The
customer or the user of the ENDOULTRA should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The ENDOULTRA uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment
RF emissions
CISPR 11
Class B
The ENDOULTRA is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies buildings
used for domestic purposes
Harmonic emissions
IEC 61000-3-2
A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
- electromagnetic immunity - for all ME
EQUIPMENT and ME SYSTEM.
electromagnetic immunity
The ENDOULTRA is intended for use in the electromagnetic environment specified below. The customer
or the user of the ENDOULTRA should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
Discharge(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for
power
supply lines
Mains power quality should be that of a
typical commercial or hospital
environment
Surge
IEC 61000-4-5
±1 kV line to line
±1 kV line to
line
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
<5%
U
T
(>95% dip in
U
T
)
For 0,5 cycle
40%
U
T
<5%
U
T
(>95% dip in
U
T
)
For 0,5 cycle
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of
ENDOULTRA requires continued
operation during power mains
EMC COMPATIBILITY
Electromagnetic compatibility (EMC) is assessed with reference to the following
standard: 60601-1-2
Guidance and manufacturer’s declaration - electromagnetic emissions - for all ME
EQUIPMENT and ME SYSTEM.
19