Distance calibration
Before starting distance calibration, is it a good idea to use the LabRail tool to check
the laser from the LabRail is aligned all the way along the rail to the LabRail Injector.
The tool window can be opened by clicking
Setup
→
Labrail
→
Tool:
Turn on the laser and use the move buttons to move the LabRail along the rail as
shown above.
Before the rail can be used, the distance between the mounted sensor and the
LabRail must be calibrated and stored in the memory of the LabRail.
Note: Fix the LabSpion goniometer to the floor so that the position does not change.
If the LabRail or the LabSpion are moved, all distance calibration procedures must be
repeated.
To make the calibration please make sure you have installed Viso Light Inspector
version 6.05.2 or later.
Go to menu
Setup
→
LabRail
and the following dialog will appear