Every time the file is renamed in
ultramic384.err
the previous configuration stored into
the microphone
will be deleted!
Scheduling Deletion
The VT UltraMic-384 stores the recording mode and ALL programming instructions into the internal memory. Even if the VT
UltraMic-384 is disconnected from the USB all these values are retained into the memory supplied by the internal watch battery.
If you want to delete the scheduling there are three ways to do it:
-
Remove the backup battery
, every
scheduling is deleted herein, included
the internal clock.
-
Use the
Android app
and once
connected click on
CONF. ERASE
.
The clock is maintained.
-
Save the following
xml file
and
store it into the VT UltraMic-384 as
previously explained for normal
scheduling:
-
Contrary to the standard scheduling
programming, the
ultramic384.xml
file will
be renamed to
ultramic384.err
and the
scheduling will be deleted.
<?xml version="1.0" encoding="UTF-8"?>
<!-- THIS FILE MUST BE SAVED AS ultramic384.xml
Configuration file to blank timer and date time
on device -->
<UltraMic384>
<config>
<defaultsettings>
<filesize>460</filesize>
<devicename>OFF</devicename>
<timeleft>1</timeleft>
</defaultsettings>
<scheduling>
</scheduling>
</config>
<setdatetime>2016-10-15T10:55:00</setdatetime>
</UltraMic384>
XML file example to