3.3.2 Dose Alarm Selection
VIQUA’s Adenovirus system comes equipped with real-time UV dose monitoring which
utilizes data from both the UV sensor (3.3.3) and the flow meter (3.3.5). The system will
go into alarm at different points depending on input from these components. There are
two dose settings to choose from: variable or stepped dose alarm. The system will
come factory set as variable but may be set to stepped depending on regional
regulations.
60
65
70
75
80
85
90
95
100
0
5
10
15
20
25
%U
V
T
System Flow (GPM)
Variable Dose Alarm
Normal operation
60
65
70
75
80
85
90
95
100
0
5
10
15
20
25
%U
V
T
System Flow (GPM)
Stepped Dose Alarm
Normal operation
System alarm
System alarm