Operator’s Manual Page 38 of 89
OXY-2-USM-1, REV K
8. Reviewing StO
2
Data
To review data, there must already be a file started by patient sampling. Tap the
Review StO
2
button to the right of the desired
channel’s StO
2
graph. The
Review StO
2
screen is displayed. Refer to Figure 11.
A graph is displayed for the amount of time elapsed from the moment sampling was started to the moment the
Review StO
2
button was
pressed. Move through the graph by using the arrows. Pressing the double arrows moves through the graph page by page (a page is the
total number of data points (90) that can be shown on one page of the graphic area). Pressing the single arrows moves through the
graph in smaller sections of data; the vertical cursor can also be positioned by tapping the screen.
The y-axis displays the percentage of StO
2
in the tissue being measured. The y-axis does not auto-adjust. The x-axis displays the data
points as they are gathered and represents the time scale. The red dotted line(s) across the graph indicates the alarm level(s) and
indicates where the data points fall with regard to the alarm level(s). The set alarm level is presented numerically to the left of the graph.
Use the
Zoom +/-
buttons to display the data in different time increments. Press the up or down arrows to select the desired time
increment. For example, pressing 50% means that the data on one page represents averaged data points in 5-minute increments; refer to
Table 4.
Table 4. Percentage of Zoom
Percentage of Zoom
Meaning of Percentage
-100%
View of 1 minute increments of data
around where cursor is focused
-50%
View of 5 minute increments of data
around where cursor is focused
-25%
View of 10 minute increments of data
around where cursor is focused
0%
View of 30 minute increments of data
around where cursor is focused
+25%
View of 1 hour increments of data
around where cursor is focused
+50%
View of 2 hour increments of data
around where cursor is focused
Summary of Contents for T.Ox OXY-2-USM-1
Page 32: ...Operator s Manual Page 32 of 89 OXY 2 USM 1 REV K Figure 7 Low Signal Quality Alarm ...
Page 37: ...Operator s Manual Page 37 of 89 OXY 2 USM 1 REV K Figure 11 Review StO2 Screen ...
Page 46: ...Operator s Manual Page 46 of 89 OXY 2 USM 1 REV K Figure 15 View Numerics Screen Single Mode ...
Page 72: ...Operator s Manual Page 72 of 89 OXY 2 USM 1 REV K Figure 33 Discard Confirmation Screen ...
Page 75: ...Operator s Manual Page 75 of 89 OXY 2 USM 1 REV K Figure 35 Reboot Recovery Screen ...
Page 76: ...Operator s Manual Page 76 of 89 OXY 2 USM 1 REV K Figure 36 Sensor Data Disposal ...
Page 84: ...Operator s Manual Page 84 of 89 OXY 2 USM 1 REV K ...